PC 3000 Multi Parameter Patient Monitor:

- 7'' high resolution colour TFT display, 5-channel waveform display
- Standard parameters: ECG, RESP, Sp02, NIBP, 2-TEMP, PR, Arrhythmia and ST segment analysis
- Up to 800 groups parameter data & 24-hour ECG waveform storage and recall
- 480-hour graphic and tubular trends for all parameters
- 7 channels ECG simultaneous waveforms display
- Protection against interference from defibrillator, electrosurgical knife
- Adult/Pediatric measurement modes
- Visual and audible alarms; networking capability

Catalog
  • Product code: 35134
  • Units: 1 pc.
  • Specifications:
    NIBP Monitoring
    • Measuring method: Automatic measurement by oscillating method
    • Cuff inflation time: <10 seconds (typical adult cuff)
    • Initial cuff inflation pressure: Adult mode: <180mmHg infant mode: <120mmHg
    • Overpressure protection limit: Adult mode: 300mmHg Infant mode: 240 mmHg

    Measuring range
    • Systolic pressure: Adult: 40-225 mmHg Infant: 40-200 mmHg
    • Diastolic pressure: Adult: 20-215 mmHg Infant: 20-165 mmHg
    • Mean arterial pressure: Adult: 10-195 mmHg Infant: 10-150 mmHg
    • Measurement accuracy: Systolic ±10% Diastolic ±10% MAP ±10%
    • Measurement mode: Manual, Auto, STAT
    • Automatic measuring: 1~240min
    • Intervals: interval tolerance: <10s

    TEMP Monitoring
    • Measuring range, accuracy: 25.0°C~45.0°C, ±0.2°C

    SpO2 & Pulse Rate Monitoring
    • Transducer: dual-wave length LED
    • SpO2 measuring range: 35%~100%
    • Low perfusion capability: 0.4%~5%
    • SpO2 display accuracy: 70%~100% ±3%
    • PR measuring range: 30bpm~250bpm
    • PR measuring accuracy: ±2bpm

    ECG Monitoring
    • Heart rate measurement: ±0.4mVp~±5mVp
    • Heart rate display range: 20bpm~250bpm
    • Display accuracy*: ±1% or ± 2 bpm, whichever is greater
    • Heart rate alarm delay time: =10s
    • Sensitivity selection: Auto, ×1/2, ×1, ×2, ×4 tolerance: =5%
    • Sweeping speed: 12.5 mm/s, 25mm/s, 50mm/s tolerance: =10%
    • ECG noise level: =30µVP-P
    • ECG input loop current: =0.1µA
    • Input impedance: =5MO
    • Common-mode rejection ratio: =89dB
    • Time constant: Monitoring mode=0.3s
    • Diagnostic mode=3.2s
    • Frequency response: Monitoring mode: 0.5Hz~25Hz
    • Diagnostic mode: 0.05Hz~75Hz

    RESP Monitoring
    • Monitoring range, accuracy: 0rpm~120rpm, ±5% or 2rpm, whichever is greater
    • RESP alarm range: 0bpm~100bpm
    • Alarm tolerance: ±10%

    Other Technical Specifications
    • Power supply: 100~250Vac, 50/60Hz
    • Power consumption: <80 VA
    • Display mode: 7 inches TFT color LCD
    • Alarming mode: Audible-visual alarm
    • Networking port: LCD serial/Net port