DIATERMO MB240 HOSPITAL
These HF electrosurgical unit can be used to execute monopolar and bipolar surgery in the many fields of application where high precision and reliability are essential, as in the endoscopic surgery.
EASY TO USE
Graphic icons allow an easy selection of the operation mode.
All operating conditions and selected output controls are represented by bright digital indications.
CUT
The monopolar cut can be performed by pure sine current to obtain an incision with a minimum damage for the adjacent tissues and sure recovery for first intention.Two different modalities of modulation allow to reduce bleeding during monopolar cut. In the bipolar modality the cut can be effected with pure sine current.
HEMOSTASIS
When the surgical action requires the obliteration of a blood vessel the availability of currents with different coagulative powers enables this type of work both with monopolar and bipolar surgery.
FULGURATION
When it is necessary to cause a superficial necrosis of wide areas, protecting the deeper tissues, a special coagulant current that dehydrates the superficial layer of the tissue and reduces the possibility of damage those beneath is available for the operation.
OUTPUT CONTROL
Output power levels are adjustable by through incremental encoders that allow the exact pre-sets and repetition of the values.The output function is linearized and controlled by ultra-modern electronic micro-controllers which will interrupt power if any alarm condition happens.
SAFETY OF USE
Circuits dedicated to the safety of operator and patient are the priority. This ensure the leakage current is kept well under the values by the international requirements, under every output condition. Monopolar and bipolar output circuits are completely floating and the use of the defibrillator is allowed while the reference plate patient is connected.
RELIABILITY
DIATERMO MB uses all solid state electronic circuits and is made with interchangeable units for a fast and effective maintenance. Automatic control circuits protect the units from damage also in case of repeated improper use.
Catalog
EASY TO USE
Graphic icons allow an easy selection of the operation mode.
All operating conditions and selected output controls are represented by bright digital indications.
CUT
The monopolar cut can be performed by pure sine current to obtain an incision with a minimum damage for the adjacent tissues and sure recovery for first intention.Two different modalities of modulation allow to reduce bleeding during monopolar cut. In the bipolar modality the cut can be effected with pure sine current.
HEMOSTASIS
When the surgical action requires the obliteration of a blood vessel the availability of currents with different coagulative powers enables this type of work both with monopolar and bipolar surgery.
FULGURATION
When it is necessary to cause a superficial necrosis of wide areas, protecting the deeper tissues, a special coagulant current that dehydrates the superficial layer of the tissue and reduces the possibility of damage those beneath is available for the operation.
OUTPUT CONTROL
Output power levels are adjustable by through incremental encoders that allow the exact pre-sets and repetition of the values.The output function is linearized and controlled by ultra-modern electronic micro-controllers which will interrupt power if any alarm condition happens.
SAFETY OF USE
Circuits dedicated to the safety of operator and patient are the priority. This ensure the leakage current is kept well under the values by the international requirements, under every output condition. Monopolar and bipolar output circuits are completely floating and the use of the defibrillator is allowed while the reference plate patient is connected.
RELIABILITY
DIATERMO MB uses all solid state electronic circuits and is made with interchangeable units for a fast and effective maintenance. Automatic control circuits protect the units from damage also in case of repeated improper use.
Catalog
- Product code: 30690
- Units: 1 pc.
- Specifications:
• Activation devices
• Bipolar: footswitch
• Dual: bipolar – footswitch, monopolar -manual
• Monopolar: footswitch - manual
• Mains: 115/230 Vac 50-60 Hz
• Absorption: 800 VA
• Class I CF
• Weight: 15 kg
• Size: 370 x 470 x h 150 mm
• Warranty: 12 months (accessories not included)
• Norms: IEC 601-2-2 - IEC 601-1 CE, 93/42/EEC
• Monopolar output - Working frequency: 475 kHz
• Cut output
• Max power: 250W
• Peak to peak voltage: 1,4 kV
• Load resistance: 300 Ohm
• Cut/coag 1 output
• Max power: 225W
• Peak to peak voltage: 1,4 kV
• Load resistance: 300 Ohm
• Cut/coag 2 output
• Max power: 200W
• Peak to peak voltage: 1,4 kV
• Load resistance: 300 Ohm
• Coag 1 output
• Max power: 150W
• Peak to peak voltage: 1,4 kV
• Load resistance: 300 Ohm
• Coag 2 output
• Max power: 100W
• Peak to peak voltage: 1,4 kV
• Load resistance: 300 Ohm
• Coag 3 output (spray):
• Max power: 80 W
• Peak to peak voltage: 4,2 kV
• Load resistance: 2000 Ohm
• Bipolar output - Working frequency: 525 kHz
• Cut output
• Max power: 90W
• Peak to peak voltage: 0,5 kV
• Load resistance: 100 Ohm
• Coag output
• Max power: 80W
• Peak to peak voltage: 0,5 kV
• Load resistance: 100 Ohm