Welcome on Sanitaria Triestina Store
  • DIATERMO MB200 - mono/bipolar - 200 W

DIATERMO MB200 - mono/bipolar - 200 W

DIATERMO MB 200 and MB 202 are HF electrosurgical units suitable to effect middle monopolar or bipolar electrosurgery. The available waveforms and peak voltage value, as well as the accurate HF leakage current operation for polypectomy, TURP, intrauterine resections, endometriosis excision, etc. The units allow the accurate preset of the output power with as low as 1 Watt resolution and to perform coagulation with high crest factor current, to get an elevated thermal effect, as well as with low voltage low crest factor current, to get coagulation in absence of carbonization. Ten different programs related to ten different operative conditions could be stored and easily recalled by the surgeon. The use of either split or normal neutral plates are allowed. A special electronic circuit (Skin Plate Electronic Control) continuously warns the plate to patient contact safety and unwanted short circuit in the connecting cable. An over time warning (OVT) is provided to prevent excessive heating of the tissue around the surgical target due to too long current delivery time. DIATERMO MB 202, furthermore, can be used with current delivery time equal to that of actuation of the output power switch as well as with preset current delivery time in the range from 1 ms up to 999 ms.
The output pulse can be automatically repeated 1 to 250 times. The presetting of the current delivery time is propre for collecting of bioptic samples and for microsurgical dermatologic operationy like treatment of telangiectasias, spider nevus, angiomas, efc. Bipolar cable and forceps are not included in standard specifications.

Catalog
  • Product code: 30542
  • Units: 1 pc.
  • Specifications:
    • Minimum preselectable power: 0
    • Level step digital: 1
    • Maximum output power CUT (W): 200 - 250 ohm
    • Maximum output power CUT 90% (W): 200 - 250 ohm
    • Maximum output power CUT 80% (W): 200 - 250 ohm
    • Maximum output power CUT 70% (W): 200 - 250 ohm
    • Maximum output power BLEND (W): 120 - 200 ohm
    • Maximum output power COAG FORCED (W): 150 - 150 ohm
    • Maximum output power COAG SOFT (W): 90 - 100 ohm
    • Maximum output power BIPOLAR (W): -
    • Maximum output power BIPOLAR COAG (W): 100 - 100 ohm
    • Maximum output power BIPOLAR CUT (W): 120 - 250 ohm
    • Crest Factor CUT: 1.5
    • Crest Factor CUT 90%: 1.6
    • Crest Factor CUT 80%: 1.7
    • Crest Factor CUT 70%: 1.8
    • Crest Factor BLEND: 2.1
    • Crest Factor COAG FORCED: 2.0
    • Crest Factor COAG SOFT: 1.7
    • Crest Factor BIPOLAR: -
    • Crest Factor BIPOLAR CUT (W): 1.5
    • Working frequency: 600 kHz
    • Weight: 8 Kg
    • Size LxHxD: 360 x 150 x 265 mm
    • Selectable mains power(Vac): 115 -230
    • Mains frequency (Hz): 50-60
    • Fuses 230Vac (5x20) TIMED: 2x 3.15A
    • Fuses 115Vac (5x20) TIMED: 2x 6.3A
    • Power accuracy output warning system
    • Split or not split patient plate allowed
    • Timed delivery repetition: -
    • Electrical Class (EN60601 - 1): I CF MDD 93/42/EEC Class: II b
    • Patient circuit: F
    • Duty cycle (action/pause) in seconds: 10/30